SMITH & NEPHEW, INC. IMAGE MANAGEMENT SYSTEM 660HD-E; DEVICE, DIGITAL IMAGE STORAGE, RADIOLOGICAL
|
Back to Search Results |
|
Catalog Number 72204668 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/08/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that during a procedure, when the right camera button was pressed to record, it was going to white balance settings instead.It was brought in a different camera head to test if that was where the default was and it was not.The image management system began taking videos without assistance one after the other.It would not come off of recording nor would it allow to cancel or close out of anything.The procedure was completed with a back up device.There was a delay greater than 30 minutes and no further complications were reported.
|
|
Manufacturer Narrative
|
H3, h6: the reported device was received for evaluation.A visual inspection did not reveal any issues.A functional evaluation was performed on the returned device and did not find any malfunction.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since the reported malfunction could not be duplicated during the investigation.Factors that can contribute to the reported event include issues during setup of the device.No containment or corrective actions are recommended at this time.
|
|
Search Alerts/Recalls
|
|
|