Trackwise#: (b)(4).The device was returned to the factory for evaluation on 02/19/2024.An investigation was conducted on 02/22/2024.A visual inspection was conducted.Signs of clinical use and evidence of blood was observed.The handle was observed to be intact with no visual defects observed.A mechanical evaluation was conducted.The blue toggle was manipulated to retract and extend the c-ring.There were no visual defects observed on the intact c-ring.The blue toggle was manipulated to retract and extend the cutter blade.There were no visual defects observed on the intact cutter blade.There were no visual defects observed on the bipolar jaws of the device.An electrical evaluation was conducted.A pre-cautery test was performed and repeated 10 times over a 10 minute period according to the procedure in the instructions for use (b)(6).The device passed the pre-cautery test with a reference generator (recommended setting of 18) and reference bipolar cord.The bipolar cord connection was manipulated during activation.The device did produce steam and the saline did ¿boil¿ on the test gauze each time.Based on the returned condition of the device, as well as the evaluation results, the reported failure "mechanical problem" was not confirmed.The lot # 3000355823 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|