• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FARAPULSE, INC. FARADRIVE STEERABLE SHEATH CLEAR; VASCULAR GUIDE-CATHETER, SINGLE-USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FARAPULSE, INC. FARADRIVE STEERABLE SHEATH CLEAR; VASCULAR GUIDE-CATHETER, SINGLE-USE Back to Search Results
Lot Number CL12230
Device Problems Leak/Splash (1354); Device Contamination with Body Fluid (2317)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/06/2024
Event Type  malfunction  
Event Description
It was reported that a faradrive steerable sheath clear during procedure for an atrial fibrillation blood was noticed flowing from the valve of the sheath, the valve was not hermetic.This happened after the physician performed the transseptal puncture with the sheath, he removed the dilator and did the purge of the material, then connected the irrigation, but a few seconds after blood was noticed.To solve the issue, the sheath was replaced because there was a risk of air introduction in the patient.Then the procedure was completed without patient complications.Device return status is unknown.
 
Manufacturer Narrative
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FARADRIVE STEERABLE SHEATH CLEAR
Type of Device
VASCULAR GUIDE-CATHETER, SINGLE-USE
Manufacturer (Section D)
FARAPULSE, INC.
3715 haven avenue
suite 110
menlo park CA 94025
Manufacturer (Section G)
CENTERPOINT SYSTEMS
3338 parkway blvd
west valley city UT 84119
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key18818495
MDR Text Key336884996
Report Number2124215-2024-11625
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Lot NumberCL12230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/06/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/30/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-