Model Number 176657 |
Device Problems
Entrapment of Device (1212); Mechanics Altered (2984)
|
Patient Problem
Tissue Breakdown (2681)
|
Event Date 02/12/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
D10 concomitant product: 176657 endoclip ii ml 10 appli (lot#j3g3176ny).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a laparoscopic video-assisted thoracic surgery (vats) procedure, once a clip was placed around the tissue and the jaws were fully squeezed, the jaws would not open and the clip remained jammed in the jaws.Both the clip applier jammed and would not release on the tissue.To resolve the issue, the surgeon had to oversew the area and cut the clip appliers free from the tissue.
|
|
Manufacturer Narrative
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that when the handle was cycled to load a clip; the clip ejected from the jaws unformed.Due to the noted damage, further functional testing of the instrument was not performed.It was reported that the instrument locked on tissue, and clips not loading properly.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if the instrument is applied over an obstruction causing the jaws to misalign and damaging the component allowing the clips to eject from the device jaws unformed or malformed.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: if firing a clip over another clip or other obstructions may result in bleeding and/or leakage and may damage the instrument jaws.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|