Model Number CI-1601-04
Device Problem Migration (4003)
Patient Problem Head Injury (1879)
Event Date 01/11/2024
Event Type  Injury  
Event Description
The recipient reportedly experienced device migration as a result of a head trauma incident.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.The recipient was successfully activated after revision surgery.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed and silicone damage was observed on the top and bottom covers of the device.This is believed to have occurred during revision surgery.Photographic imaging inspection revealed a broken antenna coil wires.This is believed to have occurred during revision surgery.System lock was lost while manipulating the antenna.This is due to the antenna condition.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical test performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Interim report was requested by the nca: the explanted device is undergoing analysis.The external visual inspection revealed the electrode was severed and silicone damage was observed on the top and bottom cover of the device.This is believed to have occurred during revision surgery.The photographic imaging inspection in pending.System lock was verified.The condition of the electrode prevented an electrical test from being performed.The device passed some the electrical tests performed.The residual gas analysis test is pending.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA 3D IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
heidi white
28515 westinghouse place
valencia, CA 91355
MDR Report Key18822187
MDR Text Key336733983
Report Number3006556115-2024-00230
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeSA
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date03/31/2023
Device Model NumberCI-1601-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2024
Is the Reporter a Health Professional? No
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/09/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age4 YR
Patient SexMale
-
-