A united states (us) customer contacted a siemens customer care center (ccc) and reported that falsely depressed sodium (na) results were obtained on three patient samples on the dimension exl with lm instrument.Quality control (qc) was in range on the day of testing.A siemens customer service engineer (cse) was dispatched to the customer¿s site.During the visit, the cse replaced both solenoids, flush pump, integrated multisensor technology (imt) port valve, tower assembly, and sample cable conduit.Then, the cse verified the tubing¿s integrity and replaced any necessary tubing, realigned the sampler to the port valve, reprimed the imt system, successfully calibrated the instrument, checked the pump rate, and ran patient samples and a precision study, which recovered acceptably.The cause of the event is related to the instrument and normal troubleshooting resolved the issue.The customer indicated that the issue initiated on (b)(6) 2024 through (b)(6) 2024.However, the customer only provided examples from (b)(6) 2024.The instrument is performing according to specifications.No further evaluation of this device is required.
|