Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).D10: cat#110018822 lot# zb7175654 g7 positioning guide rod.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated report: 0001825034-2024-00597.
 
Event Description
It was reported that while impacting the cup, the bottom of the positioning pin broke.The tip of the straight acetabular shell inserter also broke while removing the liner impactor head.There was no patient impact, no medical or surgical interventions, no surgical delay, and no foreign body retained.Surgical technique was utilized.No additional information.
 
Event Description
No further event information is available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4;g3;h2;h3;h6.The product was returned and evaluated.Medical records were not provided.A visual examination of the returned product identified that the inserter had fractured at the threaded tip.There were impact marks on the strike plate and scuffing to the stem.Review of the device history records identified no deviations or anomalies during manufacturing.Reported event is not related to a combination of products; therefore, a compatibility review is not applicable.The reported issue was confirmed based on the evaluation of the returned product; however, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR MONOBLOCK SHELL INSRTR
Type of Device
INSTRUMENT, HIPS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18825644
MDR Text Key336767281
Report Number0001825034-2024-00596
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110003450
Device Lot Number220790
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-