Device evaluation details: the product was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.A visual inspection and electrical evaluation of the returned device were performed following bwi procedures.Visual analysis indicated that the pebax material was found in an inflated condition with foreign matter present.While the root cause of this condition could potentially be related to the usage of the device during the procedure, however, it cannot be conclusively determined.The issue is unrelated to the reported event.Additionally, an electrical test was conducted, however, it failed due to an open circuit identified in the tip area.Subsequent inspection revealed a hole in the pebax, which exposed internal components.The presence of foreign matter could possibly be attributed to the hole, which may have entered through the hole, causing this condition.However, this cannot be determined.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.The instructions for use contain the following recommendation: ecg noise is typically generated as the result of the improper connection of the body surface ecg patch to the patient.This noise is the most significant during ablation.To resolve this situation, verify the proper connection.It is recommended to turn off the notch filter for this verification.A manufacturing record evaluation was performed for the finished device, and no internal action was found during the review.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.Explanation of codes: investigation findings: open circuit (c0205) / investigation conclusions: cause not established (d15) / component code: electrical lead/wire (g02015) were selected as related to open circuit issue.Investigation findings: mechanical problem identified (c07) / investigation conclusions: unintended use error caused or contributed to event (d1102) / component code: sleeve (g04115) were selected as related to the hole in the pebax.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|