HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP; VENTRICULAR (ASSISST) BYPASS
|
Back to Search Results |
|
Model Number 1104 |
Device Problems
Obstruction of Flow (2423); Pumping Problem (3016)
|
Patient Problems
Dyspnea (1816); High Blood Pressure/ Hypertension (1908); Obstruction/Occlusion (2422); Thrombosis/Thrombus (4440); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/27/2024 |
Event Type
Injury
|
Event Description
|
It was reported that the ventricular assist device (vad) exhibited low flow alarms. the patient was previously hospitalized for respiratory failure and high blood pressure which was treated with medication and the patient was discharged. the patient came to the hospital when the low flow alarms were more frequent. right ventricular failure was suspected as the tricuspid valve and mitral valve have regurgitation stage 3 but there were no issues found with the right ventricle. there was no change when the speed was increased, the septum remains midline. the patient had a computerized tomography (ct) with contrast done to assess outflow graft. the vad remains in use. no further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted as additional information has being received for this event and sections have been updated.Additional product: d1: heartware ventricular assist system ¿ outflow graft d4: model #: 1125 catalog #: 1125/ expiration date: unk / serial or lot#: unk udi #:ask d9: no h4:unk h5: yes h6: the codes present in section h6 correspond to components/products that comprise the reported event.Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was further reported that the patient received treatment at an external center and returned to normal after medication for high blood pressure.The patient was admitted to the hospital due to shortness of breath.The patient received medication and was placed on a ventilator.A computed tomographic angiography (cta) revealed occlusion in the outflow graft at the anastomosis level.It was thought to be right insufficiency, but it turned out to be graft occlusion.An outflow graft stenting was performed.The occlusion did not cause any problems in hemodynamics.The vad parameters returned to the normal values.No neurological or occlusive vascular injury were detected.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for device lot number./ serial or lot#: (b)(6).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|