The investigation determined that lower than expected vitros nbnp2 results were obtained when processing a non-vitros quality control fluid on a vitros 5600 integrated system.The cause of the lower than expected results was due to an issue with the vitros 5600 integrated system.The historical vitros nbnp2 quality control results were acceptable in regard to both accuracy and within laboratory precision prior to the event.In addition, the quality control results for vitros nbnp2 lot 0200 were acceptable on the other vitros 5600 integrated system in the customer's laboratory when using the same quality control fluids ruling out an assay issue as the cause of the event.The ortho field engineer (fe) performed service actions on the vitros 5600 integrated system including replacing the waste connectors from the fluid manifold to the waste cap, all three microwell incubator driver motors, each incubator ring wear pads and the well wash linear motors and the nozzles.The ortho fe cleaned and rebuilt the pre and final well wash assemblies and performed all required adjustments.A full luminometer calibration was performed, and all required assay calibrations were performed.Once service actions were completed, the ortho fe attempted to perform post service within-run precision testing, however, the precision testing could not be completed due to mechanical failures.Further service actions are required to mitigate the instrument performance issues.The ortho fe will return to the customer site to complete all service actions which are expected to return the vitros 5600 system to the intended performance.
|