BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number D134805
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Paralysis (1997)
Event Date 02/08/2024
Event Type  Injury  
Event Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and the patient experienced diaphragmatic nerve paralysis.Diaphragmatic nerve paralysis occurred the day after the procedure day around midnight after the procedure.The patient is under observation.The ablation was performed for superior vena cava (svc) isolation at 35 watts and ablation index (ai) as a little below 400.The lateral side was ablated after confirming twitching before ablation.Additional information was later received indicating the patient vomited, x-ray examination revealed elevation of the right diaphragm.Physician's judgment on health hazard is that it was non-serious (moderate/minor).The adverse event was considered to be related to the product and procedure.It seemed to be the cause of the phrenic nerve paralysis that the ablation was performed at the proximal side of the twitching site with high contact force (cf) and ai although checking the twitching site prior to the ablation.The patient did not require extended hospitalization.No intervention was provided.Patient reported as ¿fully recovered¿ and has been discharged.
 
Manufacturer Narrative
Device investigation details: the device has been reported as discarded, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.A manufacturing record evaluation was performed for the finished device number lot 31173676l and no internal action related to the complaint was found during the review.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.E1.Initial reporter phone: (b)(6).If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18829269
MDR Text Key336861353
Report Number2029046-2024-00713
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835010183
UDI-Public10846835010183
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberD134805
Device Lot Number31173676L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/09/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTO 3 SYSTEM; SMARTABLATE GEN. KIT (JAPAN)
Patient Outcome(s) Other;
-
-