• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801188
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2024
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) verified that the oxygen (o2) sensor hours were exceeded per the electronic patient gas system (epgs) log.The epgs was replaced.The unit operated to the manufacturer's specifications.Per data log analysis, it was found that the date changed on the ccm, causing the year to appear to be much later than actual which caused the 'service gas system' message to appear.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the central control monitor (ccm) displayed a 'service gas system' message.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The service repair technician (srt) could not duplicate the reported complaint.The electronic patient gas system (epgs) was powered on and air and oxygen (o2) were introduced.The o2 analyzer calibrated successfully and all testing passed.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key18830875
MDR Text Key336875963
Report Number1828100-2024-00073
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799000588
UDI-Public(01)00886799000588(11)201009
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801188
Device Catalogue Number801188
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/09/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-