TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 801188 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/08/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The field service representative (fsr) verified that the oxygen (o2) sensor hours were exceeded per the electronic patient gas system (epgs) log.The epgs was replaced.The unit operated to the manufacturer's specifications.Per data log analysis, it was found that the date changed on the ccm, causing the year to appear to be much later than actual which caused the 'service gas system' message to appear.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the central control monitor (ccm) displayed a 'service gas system' message.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The service repair technician (srt) could not duplicate the reported complaint.The electronic patient gas system (epgs) was powered on and air and oxygen (o2) were introduced.The o2 analyzer calibrated successfully and all testing passed.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|