Model Number WA53000A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/19/2024
Event Type  malfunction  
Manufacturer Narrative
The device was returned and the evaluation found that there was a distorted image caused by damages inside the optical system.The lens as initially reported was not broken.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported, that the laparoscope had its lens broken.The issue was found during preparation for use.There were no reports of patient harm.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the lens broken was not confirmed.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 10 MM, 0°, HD, QUICK LOCK, AUTOCLAVABLE
Type of Device
LAPAROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18830997
MDR Text Key337752570
Report Number9610773-2024-00608
Device Sequence Number1
Product Code EWY
UDI-Device Identifier04042761052757
UDI-Public04042761052757
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K950076
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA53000A
Device Lot Number625268
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/02/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-