DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE D PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-33
Device Problem Failure to Zero (1683)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2024
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported during a routine check performed by the customer, the cardiosave intra-aortic balloon pump (iabp) user was unable to perform zeroing on the ibp line.There was no patient involvement.
 
Manufacturer Narrative
Updated fields: b4, d9, g3, g6, h2, h3, h4, h6, h10.A getinge field service engineer (fse) was dispatched to evaluate the unit.The (fse) reported that the issue was user error.The end user did not properly use the screen, and the unit was unable to be zeroed.The unit passed all functional and safety tests and was returned to customer in working condition.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE D PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18832031
MDR Text Key336876178
Report Number2249723-2024-00917
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567109008
UDI-Public10607567109008
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-33
Device Catalogue Number0998-00-0800-33
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/23/2024
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-