STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN KNEE; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
Catalog Number UNK_JR |
Device Problem
Mechanical Problem (1384)
|
Patient Problems
Pain (1994); Loss of Range of Motion (2032)
|
Event Date 02/08/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
Reported event: an event regarding range of motion (rom) involving an unknown knee was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.-clinician review: no medical records were received for review with a clinical consultant.-product history review: could not be performed as the device lot details were not provided.-complaint history review: could not be performed as the device lot details were not provided.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
|
Event Description
|
It was reported through the stryker facebook page, "even there & done that; 2002 & 2006.No more pain, but still hard to climb up stairs; not down.".
|
|
Search Alerts/Recalls
|
|
|