• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN KNEE; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN KNEE; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number UNK_JR
Device Problem Mechanical Problem (1384)
Patient Problems Pain (1994); Loss of Range of Motion (2032)
Event Date 02/08/2024
Event Type  Injury  
Manufacturer Narrative
Reported event: an event regarding range of motion (rom) involving an unknown knee was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.-clinician review: no medical records were received for review with a clinical consultant.-product history review: could not be performed as the device lot details were not provided.-complaint history review: could not be performed as the device lot details were not provided.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
It was reported through the stryker facebook page, "even there & done that; 2002 & 2006.No more pain, but still hard to climb up stairs; not down.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN KNEE
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
juana hiciano
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18836678
MDR Text Key336890054
Report Number0002249697-2024-00343
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/05/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK_JR
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received02/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-