MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRELINK; DATA MANAGEMENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRELINK; DATA MANAGEMENT SYSTEM Back to Search Results
Model Number 71732-01
Device Problem Device Alarm System (1012)
Patient Problems Fatigue (1849); Hypoglycemia (1912); Loss of consciousness (2418); Convulsion/Seizure (4406)
Event Date 02/14/2024
Event Type  Injury  
Manufacturer Narrative
The most probable root causes associated with this failure mode are software/data corruption or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The customer reported a missing high/glucose alarms with the freestyle librelink application.Abbott diabetes care attempted to replicate the reported issue and the reported configuration of the device and android 14 is not compatible with the freestyle librelink application.The compatibility guide is available to the customer on the abbott diabetes care website.As the compatibility guide is provided to the customer, and the incompatible configurations were used, this complaint is therefore not confirmed to use.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A customer reported an alarm issue with the adc device in use with a google pixel 7 phone with android version 14.The customer indicated that the high/low glucose alarm did not sound and was therefore not made aware of changing glucose levels.The customer experienced tiredness, seizure, and loss of consciousness, requiring treatment of intravenous glucose.There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRELINK
Type of Device
DATA MANAGEMENT SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key18842331
MDR Text Key336956191
Report Number2954323-2024-07742
Device Sequence Number1
Product Code PZE
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71732-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/15/2024
Initial Date FDA Received03/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/15/2024
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-