• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COCHLEAR LIMITED NUCLEUS CI522 COCHLEAR IMPLANT WITH SLIM STRAIGHT ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR LIMITED NUCLEUS CI522 COCHLEAR IMPLANT WITH SLIM STRAIGHT ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number CI522
Device Problem Positioning Failure (1158)
Patient Problem Pain (1994)
Event Type  Injury  
Manufacturer Narrative
This report is submitted on march 06, 2024.
 
Event Description
Per the clinic, the patient experienced pain and poor performance with the device.Reprogramming attempts were made; however, the issue could not be resolved.The tip of the electrode array also was found to be folded over in the cochlea, resulting in the decision to explant the device.The device was explanted on (b)(6) 2023, and patient was reimplanted with another device during the same surgery.
 
Manufacturer Narrative
Device analysis report attached.This report is submitted on april 12, 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI522 COCHLEAR IMPLANT WITH SLIM STRAIGHT ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LIMITED
1 university avenue
macquarie university
nsw 2109
AS  2109
Manufacturer (Section G)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS   2109
Manufacturer Contact
amirah rosly
unit ug-1, vertical podium
no. 8 jalan kerinchi
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key18843390
MDR Text Key336955857
Report Number6000034-2024-00726
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502032599
UDI-Public(01)09321502032599(11)160408(17)180407
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date04/07/2018
Device Model NumberCI522
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2024
Is the Reporter a Health Professional? No
Date Manufacturer Received03/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-