• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SINGAPORE OPERATIONS AZURE XT DR MRI SURESCAN; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SINGAPORE OPERATIONS AZURE XT DR MRI SURESCAN; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number W1DR01
Device Problems High impedance (1291); Mechanical Problem (1384); Misconnection (1399)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2024
Event Type  malfunction  
Event Description
It was reported that during an implantable pulse generator (ipg) change out, the replacement ipg had a connector port and set screw issue with undefined high impedance observed on the right ventricular (rv) lead after multiple attempts.The ipg was attempted not used, removed and replaced.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZURE XT DR MRI SURESCAN
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SN  486056
Manufacturer (Section G)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SN   486056
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18844521
MDR Text Key336961781
Report Number3008973940-2024-01614
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier00763000316846
UDI-Public00763000316846
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberW1DR01
Device Catalogue NumberW1DR01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
-
-