The customer reported that the monitor did not alarm properly; the red alarm did not activate only yellow alarm although patient went into ventricular tachycardia (v-tach).The device was in use on patient at time of event, there was no adverse event reported.A philips remote service engineer (rse) spoke with the customer and the customer alleged that the monitor did not alarm properly; the red alarm did not activate only yellow alarm although patient went into v-tach¿.The rse informed that the customer needed configuration password for mx40.The rse gave the customer passwords for mx40's only to resolve the reported issue.A good faith effort (gfe) was made to obtain additional information and resolution for the case.In response, it was noted that the alarm issue did not cause any patient harm and configuration password resolved the reported issue.Based on the information available, the cause of the reported problem was due to the configuration issue of the device.The reported problem was confirmed.The engineer provided the configuration password to the customer to resolve the reported issue.
|