The coaguchek xs meter serial number was (b)(6).Per product labeling, error 4 indicates the test strip is unusable and to test using a new test strip.The meter and strips have been requested for return.The product has not been received at this time.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods." e3 - occupation was patient/consumer.
|
There was an allegation of error 4 and questionable results from the coaguchek xs meter.At 11:00 am, the result from the meter was 3.2 inr.At 1:00 pm, the result from a laboratory using an unknown reagent was 4.2 inr.The therapeutic range was 2.5 inr-3.5 inr.The testing frequency was not provided.
|