The event involved a 211 cm (83") ext set w/4 gang stopcock, pole mount, 8 microclave® clear, 2 check valves, 4-way stopcock, rotating luer.The event happened at the intensive care unit.The installation of the infusion occurred at 12 a.M.At 8 p.M., the nurse noted the presence of air in the tubing between the manifold and the injection site.After checking, the nurse noted the presence of a leak at the output of the manifold and notes that the tubing is disconnecting.The manifold was not manipulated with pliers or forced manually.The drugs infused were olimel n12, glucose and insulin.The status of the product at the time of the event is during infusion.There was patient involvement.There was no impact on the patient's state of health.No adverse events, no human harm, no delay in therapy, no need of medical intervention.No visible default on the device before use.No cut, no tear, no hole on the device.The treatment has been resumed and medications fully administered.No need of medical intervention for the nurse.
|
Received one used 011-mc330273 extension set for inspection.No damages, anomalies, or separations noted.Examination of the male luer of the manifold to the female luer of the tubing under uv light showed sufficient solvent coverage.The set was leak tested per product specifications.There was no leakage.The reported complaint of leakage could not be replicated or confirmed.A device history report (dhr) lot # review could not be conducted because no lot number(s) was/were identified.
|