BOSTON SCIENTIFIC DE COSTA RICA S.R.L. POLARX; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Lot Number 0032934927 |
Device Problem
Infusion or Flow Problem (2964)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/11/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Polarx fit balloon cath st was evaluated by boston scientific.Visual inspection noted that there was visible blood was seen inside the balloon and no external damage was noted.The device was not assessed with an isaac pressure decay testing system to determine if any potential leak paths existed that may have contributed to the blood detection error in the field, as there was visible blood inside the balloon region.The catheter handle was opened to further investigate the breach leading to the blood inside the balloon.The outer balloon line located inside of the handle was pressurized to locate a leak path.A leak path was found in the outer balloon itself.The reported allegation of a blood detection error was confirmed.
|
|
Event Description
|
Reportable based on analysis completed on 14feb2024.It was reported that during preparation for an atrial fibrillation procedure a crbs polarx fit st was selected for use.After connecting the catheter to the console, a blood detection circuit error occurred.High refrigerant flow message appeared briefly, but messaged disappeared and did not affected procedure.The catheter was replaced and the error was cleared.The procedure was completed successfully.No patient complications were reported.The device is expected to return for laboratory analysis.However, analysis of the returned device revealed visible blood ingress into the balloon.
|
|
Search Alerts/Recalls
|
|
|