SMITHS MEDICAL ASD, INC. LEVEL 1 LOW FLOW - HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problems Defective Alarm (1014); Fluid/Blood Leak (1250)
Patient Problem Insufficient Information (4580)
Event Date 11/01/2023
Event Type  malfunction  
Event Description
It was reported that the device leaked and the alarms failed.There was unknown patient involvement and unknown patient harm/adverse event reported.
 
Manufacturer Narrative
B3.Date of event: unknown.No information has been provided to date.H3.Reason device not evaluated by mfg: other; device was not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
H6.Evaluation codes: updated.Device investigation: no device was received for investigation.Therefore, we are unable to confirm the reported problem.If we receive the device, we will reopen the investigation for further evaluation.The most probable causes would be: a cracked tank cover and reservoir near the drain tube fitting resulting in a leak, outdated and damaged pcb board resulting in alarm failures, or broken bracket from drop damage.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.H4 - device mfg date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 LOW FLOW - HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18851475
MDR Text Key337150297
Report Number3012307300-2024-01177
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90-DC
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/06/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-