• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORAX MEDICAL, INC. 1.5T LINX, 13B; ANTI-REFLUX IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORAX MEDICAL, INC. 1.5T LINX, 13B; ANTI-REFLUX IMPLANT Back to Search Results
Catalog Number LXMC13
Device Problems Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Pyrosis/Heartburn (1883)
Event Date 02/29/2024
Event Type  Injury  
Event Description
It was reported that a patient has a defective linx and recurrent gerd symptoms.Was implanted in (b)(6) hospital on (b)(6) 2017.It will be removed and replaced with another linx.
 
Manufacturer Narrative
(b)(4).B3: unknown; captured as awareness date.Date sent: 3/6/2024.Photo images were received.Any additional information obtained from the photo evaluation will be included as part of the product investigation.No lot number was provided therefore a device history could not be done.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: when did the reflux begin? what is the device lot number? was the device initially effective in controlling reflux? were any events associated with the onset of symptoms (vomiting, retching, trauma, surgery)? did the patient undergo an mri since device implant? if so, when was the mri and what strength? did the patient have any other surgeries in the area? was any additional imaging performed since device implant? does the device appear to be in a continuous annular state in these images? we are interested in establishing a window when the device may have become discontinuous.Please share any additional images.What is the management plan? is device removal scheduled? is a replacement linx or fundoplication planned? when and if the explanation takes place can we ask that the procedure gets video recorded and the video shared? when and if the linx device is removed, may we ask that the device be returned for analysis? was ph testing performed prior to explant to confirm recurrent reflux? after implant, was the device initially effective in controlling reflux? additional information received: patient date of birth: on (b)(6) 1959, general health: great, allergies: no known allergies, any mri scan since implant: none, date of return of symptoms: < 1 year.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
(b)(4).Date sent: 3/12/2024.Photo analysis: an x-ray image of the device in vivo was reviewed by a medical safety officer.As per medical safety officer: "the image review demonstrates a discontinuous linx device." the mechanism/cause of failure is unknown as a hands-on analysis of the device is necessary to determine the cause of failure.No further investigation can be completed at this point.
 
Manufacturer Narrative
(b)(4).Additional information received: surgeon provided explant date of (b)(6) 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5T LINX, 13B
Type of Device
ANTI-REFLUX IMPLANT
Manufacturer (Section D)
TORAX MEDICAL, INC.
4545 creek road
cincinnati OH
Manufacturer Contact
kate karberg
4545 creek rd.
cincinnati, OH 45242
3035526892
MDR Report Key18851969
MDR Text Key337071697
Report Number3008766073-2024-00041
Device Sequence Number1
Product Code LEI
UDI-Device Identifier00855106005332
UDI-Public00855106005332
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberLXMC13
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age64 YR
Patient SexFemale
-
-