As reported by a field clinical specialist (fcs), during a tavr procedure with a 29mm sapien 3 ultra resilia valve in the aortic position, the delivery system balloon burst at the end of inflation.The physicians obtained bilateral groin access, right arterial, left arterial and left venous without issue.Right transfemoral (tf) access was obtained for the esheath+.Pigtail and temp pacer placed on the left side.The 29mm s3ur commander delivery system (ds) entered into the esheath+ without issue.Achieved native valve crossing without issue.During valve deployment, the ds balloon burst occurred at the end of inflation.The team paused to view echo imaging and felt there was no need for post dilation.The team proceeded forward with trying to remove the ds balloon.Partial balloon was pulled back into the esheath+ and resistance was felt.The operator pulled the delivery system into the esheath+ and noticed kinking in the sheath, however continued to pull the system with resistance.The ds balloon was very challenging to remove, and the operators were unable to fully pull the balloon into the sheath.The team pulled the sheath and ds out of the vessel, but the balloon tip and torn portion remained in the vessel at the access point.The vascular team was called for cutdown and vessel repair.The access site required cutdown due to removal of balloon.The patient required 3 units of blood due to the bleeding.The perceived root cause of the vessel tear was the delivery system balloon.The valve is functioning well, and the patient is stable.
|