Model Number 16-02-85
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/06/2024
Event Type  malfunction  
Event Description
Livanova deutschland received a report that a heater-cooler system 3t displayed alarm on patient side during a procedure.The alarm was silenced and procedure completed.There is no report of any patient injury.
 
Manufacturer Narrative
A.1.-a.5.Patient information was not provided.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in usa.There is no ¿error code¿ for this issue and is a light that looks like the outline of a person that illuminates when the patient side temperature is outside of the limit.The livanova field service representative in charge calibrated the patient safety limit and upper and lower temperature limits.Let run at the upper temperature for an hour to test - no failures.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER/COOLER 3T 110V/60 HZ
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich CO 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18854437
MDR Text Key337688780
Report Number9611109-2024-00121
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/06/2024
Date Device Manufactured09/09/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-