Catalog Number NSLX137C
Device Problems Difficult to Open or Close (2921); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2024
Event Type  malfunction  
Event Description
It was reported that, during an ob-gyn procedure, the jaws could not open after the device was used on tissue.Gen11 was used.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Manufacturer Narrative
(b)(4) date sent: 3/7/2024 d4 batch #: unknown additional information was obtained: the handle was manually returned, the jaws were opened, and the device was removed from the tissue.Some oozing was observed, but the bleeding was stopped by cauterization with an electrocautery scalpel.The patient is stable and he/she is discharged from the hospital.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
(b)(4).Date sent: 3/22/2024.D4 batch #: a9d21j.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was received with no apparent damage.The device was tested on the generator and passed all functional testing.The energy output delivered from the device was verified.All tones were heard during functional testing.The device was tested with the test media and no anomalies were found.There were no anomalies noted with the functionality of the device.The device opens or closes as expected.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device batch a9d21j, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL X1 CURVED 37CM SHAFT
Type of Device
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key18855871
MDR Text Key337104729
Report Number3005075853-2024-01976
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10705036015468
UDI-Public10705036015468
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K172580
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberNSLX137C
Device Lot NumberA9D21J
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received03/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/24/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GENERATOR
-
-