• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES DR TRUWAVE DISPOSABLE PRESSURE TRANSDUCER; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES DR TRUWAVE DISPOSABLE PRESSURE TRANSDUCER; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number PX12N
Device Problem Fluid/Blood Leak (1250)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 02/11/2024
Event Type  malfunction  
Event Description
As reported, when taking a blood sample in a neo natal patient with this disposable pressure transducer connected to an arterial line, lot of air came into the syringe and reasonable but unknown blood quantity leaked.Then, it was confirmed there were no loose connections and when flushing the system a leakage was found through the stopcock.Patient demographic unable to be obtained.No further information is available.There was no allegation of patient injury.The device was available for evaluation.
 
Manufacturer Narrative
The product is expected to be returned for analysis; however, it has not yet been received.Upon the return of the product a supplemental report will be sent with the investigation results.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Manufacturer Narrative
The device involved in this complaint was not available for evaluation.Without return of the product, edwards is unable to perform a complete investigation of the reported event.It is not possible to determine what factors may have contributed to it, and therefore no actions could be planned.Since the affected unit was not returned for evaluation, a product non-conformance or device failure could not be confirmed.As such, based on available information, a definite root cause is unable to be determined at this time.The manufacturing records were reviewed and there is no indication of a related nonconformance; all process parameters were met and inspections passed successfully.Additionally, there are manufacturing controls as leak test, manuafacturing continuous monitoring sampling or qa sampling inspection to avoid potential causes related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUWAVE DISPOSABLE PRESSURE TRANSDUCER
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer (Section G)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer Contact
samantha eveleigh
1 edwards way
irvine, CA 92614
9492503939
MDR Report Key18857141
MDR Text Key337120846
Report Number2015691-2024-01682
Device Sequence Number1
Product Code DXO
UDI-Device Identifier07460691950795
UDI-Public(01)07460691950795(17)241219(11)221220(10)64687004
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K222216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPX12N
Device Lot Number64687004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/11/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-