Catalog Number 1013-00-901
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2024
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi - (b)(4).
 
Event Description
It was reported that during an unspecified surgical procedure it was observed that the head of the impactor device broke into 2 pieces.It was not reported if there was a delay in the procedure due to the event.It was not reported if a spare device was available for use.There was patient involvement reported.There were no reports of injuries, medical intervention, or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.Quality engineering evaluated the impactor device and the reported condition that the head of the impactor broke was confirmed.An assessment was performed, and it was found that device failed visual inspection as the end cap was cracked.The cap was functionally assessed and determined to be cracked, consistent with having been dropped or misaligned during use.It was further determined that the device failed pretest for visual assessment and end cap thread assessment.The assignable root cause of this condition was determined to be traced improper handling, which is user error.A review of the device history was performed and no non-conformances were detected related to the reported condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KINCISE¿ FEM HEAD IMPACTOR, REPLACE CAP
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
TECHTRONIC INDUSTRIES MEDICAL
1428 pearman dairy rd
anderson SC 29625
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
3035526892
MDR Report Key18858145
MDR Text Key337131247
Report Number1045834-2024-00246
Device Sequence Number1
Product Code GEY
UDI-Device Identifier00845384018629
UDI-Public00845384018629
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1013-00-901
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received04/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/12/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-