Catalog Number 261221
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2024
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A physician reported a codman perforator (id 261221) disassembled when pulling the product from cranium after making the first burr hole.All the disassembled parts were recovered and the device was used to complete the surgery.A total number of burr hole made by the product was one and the drill used with the perforator was aesculap/elan 4 electro.There was less than 30 minutes delay in the procedure.The procedure performed was craniotomy for chronic subdural hematoma.According to information provided, it is unknown if the recommended spring tests were being performed between each burr hole.It is also unknown if the perforator clicked in place in the drill.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18858306
MDR Text Key337699798
Report Number3014334038-2024-00049
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number261221
Device Lot Number7263655
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/22/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
AESCULAP/ELAN 4 ELECTRO.
-
-