SYNTHES GMBH COLIBRI II HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.101
Device Problems Output below Specifications (3004); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2023
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the moving parts of the trigger not moving smoothly, identified during service and evaluation was confirmed.The assignable root cause was determined to be traced to maintenance, which is improper maintenance.Udi: (b)(4).
 
Event Description
It was reported by germany that during service and evaluation, it was determined that the moving parts of the trigger did not move smoothly on the small battery drive device.It was further observed that the device would not run because the motor was damaged.It was determined that the device had cosmetic damage due to a worn coupling.It was further determined that the device failed pretest for check the attachment coupling, check for sticky triggers, check function of device and check power with power test bench.It was noted in the service order that the device was idling.It was further reported that the device was turning in slow motion.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COLIBRI II HANDPIECE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key18861176
MDR Text Key337572657
Report Number8030965-2024-03375
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2023
Date Manufacturer Received03/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-