SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Power Problem (3010)
Patient Problem Insufficient Information (4580)
Event Date 02/01/2024
Event Type  malfunction  
Manufacturer Narrative
B3.Date of event: unknown.No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the power turned on and off, and the behavior was inconsistent.Also, the coating near the outlet was peeling off.There was unknown patient involvement and unknown patient harm/adverse event reported.
 
Manufacturer Narrative
Evaluation codes: updated.Device evaluation: one device was received.Per visual inspection, the rubber feet on the bottom of the unit had melted and ¿disappeared¿.It was also noted that there were cracks in the main cabinet.The complaint was confirmed through functional testing as the power does not turn on and off.The root cause was the ac power cord was broken due to mechanical shock such as a drop.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.At the customer's request, the product was returned without repair.D3 - address.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18861285
MDR Text Key337558035
Report Number2183161-2024-00184
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public(01)50695085820909(11)081009
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberHL-90JP
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/09/2024
Is the Reporter a Health Professional? No
Date Manufacturer Received03/17/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/09/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-