• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number CA17K0/000/004JP
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/31/2024
Event Type  malfunction  
Manufacturer Narrative
Expiration date and manufacture date are unknown, no information is available based on the reported lot number.510k is blank, device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the connection was bent while checking.There was no patient involvement and no patient harm/adverse event reported.
 
Manufacturer Narrative
Evaluation codes: updated.Device evaluation: six photos were included for evaluation; damaged connector was observed.One device component was returned for investigation.Visual observation of on the gas sampling line (item number 225-3422-800, lot number 4419607) revealed that one of its connectors was damaged (deformed).No other anomaly was observed.The complaint was confirmed.Root cause cannot be determined since the reported failure mode is not attributable to the manufacturing process.A review of the device history record (dhr) for the received component (lot 4419607) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.No action taken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ANESTHESIA BREATHING CIRCUIT
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
3350 granada ave n, suite 100
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18861335
MDR Text Key337209069
Report Number2183161-2024-00186
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCA17K0/000/004JP
Device Lot Number240111
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/15/2024
Date Manufacturer Received04/05/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-