MAUDE Adverse Event Report: WILSON-COOK MEDICAL INC MEMORY II DOUBLE LUMEN EXTRACTION BASKET; LQR DISLODGER, STONE, BILIARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILSON-COOK MEDICAL INC MEMORY II DOUBLE LUMEN EXTRACTION BASKET; LQR DISLODGER, STONE, BILIARY Back to Search Results
Catalog Number MB-35-2X4-8
Device Problem Difficult to Fold, Unfold or Collapse (1254)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2024
Event Type  malfunction  
Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use states: "confirm desired position of the basket sheath relative to the target.Advance the basket out of sheath by pushing forward on the handle.Caution: pulling on the sheath while advancing or retracting the basket may damage the device, rendering it inoperable." the instructions for use also indicates: "advance the device through the channel, in short increments, until the basket sheath exits the endoscope." basket deployment difficulties and buckling/breaking of the drive wire can occur if the device experiences excessive pressure.Resistance in basket movement and bends in the catheter can occur if the elevator of the endoscope is used to deflect the device at a sharp angle.Prior to distribution, all memory ii double lumen extraction baskets are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During removal of common bile duct stones, the physician used a cook memory ii double lumen extraction basket.It was reported that the handle of the basket did not work as they removed some common bile duct stones from the basket [basket will not advance or retract].A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEMORY II DOUBLE LUMEN EXTRACTION BASKET
Type of Device
LQR DISLODGER, STONE, BILIARY
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key18861401
MDR Text Key337180938
Report Number1037905-2024-00136
Device Sequence Number1
Product Code LQR
UDI-Device Identifier10827002251493
UDI-Public(01)10827002251493(17)261024(10)W4777297
Combination Product (y/n)N
PMA/PMN Number
K171969
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMB-35-2X4-8
Device Lot NumberW4777297
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/14/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ENDOSCOPE-OLYMPUS TJF260V
-
-