Manufacturer's investigation conclusion: a direct correlation between heartmate ii left ventricular assist system (lvas), serial number (b)(6) and the reported event of gastrointestinal bleeding could not be conclusively established through this evaluation.Review of the submitted log files confirmed silenced driveline fault alarms.Based on previous experienced with the heartmate ii lvas and similar reported events, these events could be indicative of an issue with the driveline.The submitted log files collectively contained events from 27feb2024 through 14mar2024.Several silenced driveline fault alarms were captured.No other notable alarms or events were captured.The pump appeared to function as intended at the fixed speed.The patient remains ongoing on heartmate ii lvas, serial number (b)(6) with no further complaints reported at this time.The relevant sections of the device history records for (b)(6) and the driveline, serial number (b)(6) were reviewed and showed no deviations from manufacturing or quality assurance specifications.The heartmate ii left ventricular assist system (lvas) instructions for use (ifu) rev.C, and the heartmate ii lvas patient handbook rev.C are currently available.Section 1 of the ifu, ¿introduction¿, lists potential adverse events that may be associated with the use of the heartmate ii lvas, including bleeding.Section 6 of the ifu, ¿patient care and management¿ provides information regarding anticoagulation, including the recommended inr (international normalized ratio) values.This section also discusses damage due to wear and fatigue of the driveline.This section also contains information on ¿caring for the driveline¿ (under ¿ongoing system assessment and care¿) and provides possible indications of driveline damage as well as how to respond to such events.This section also instructs to check the driveline daily for signs of damage, such as cuts, holes, or tears.Section 8 ¿equipment storage and care¿ also contains information on ¿care of the driveline,¿ and provides possible indications of driveline damage.Section 7 of the ifu, "alarms and troubleshooting" outlines system controller alarms as well as how to respond to each alarm condition.The patient handbook contains a section on ¿caring for the driveline¿; however, all heartmate ii lvad drivelines have the potential for wire/shield breakdown to occur dependent upon length of use and patient handling.Section 4 of the handbook, ¿living with the heartmate ii¿ contains information on caring for the driveline.Section 5 of the handbook, "alarms and troubleshooting" outlines system controller alarms as well as how to respond to each alarm condition.No further information was provided.The manufacturer is closing the file on this event.
|