• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FINELINE II EZ STEROX; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FINELINE II EZ STEROX; IMPLANTABLE LEAD Back to Search Results
Model Number 4471
Device Problems Mechanical Problem (1384); Defective Device (2588); Contamination /Decontamination Problem (2895); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/13/2024
Event Type  malfunction  
Event Description
It was reported that this right ventricular (rv) was not implanted successfully due to the high pacing threshold, the lead was found to have blood in the insulation, and the coil appeared to be extended.This lead was never in service, and a new lead was placed.No adverse patient effects were reported.
 
Manufacturer Narrative
The product has been received for analysis.Upon completion of the failure analysis of the complaint product, if there is any further relevant information from that review, a supplemental report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FINELINE II EZ STEROX
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18862185
MDR Text Key337556915
Report Number2124215-2024-14245
Device Sequence Number1
Product Code DTB
Combination Product (y/n)Y
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4471
Device Catalogue Number4471
Device Lot Number586703
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/09/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
-
-