During a pulmonary vein isolation for atrial fibrillation, after isolation, when the catheter was removed from the heart, it was noted that a substance like myocardial tissue was attached to the tip of the catheter.It was noted that during rf delivery there was resistance when dragging.The procedure was completed without replacing the catheter.
|
One bi-directional, curve d-f, tactiflex ablation catheter, sensor enabled was received for evaluation.One image was returned for review.The image appears to show a foreign material on the distal tip of the catheter.During inspection of the returned device, a foreign material consistent with cardiac tissue was noted on the distal tip, consistent with the reported event.The device met specifications during temperature testing, occlusion testing, leak testing, and flow rate testing.When the returned device was connected to the tactisys quartz unit, optical fibers 1-3 met specifications for optical properties and contact force was displayed with no error messages noted.Both thermocouples, the tip electrode, and the magnetic sensor met specifications during electrical testing with no open or short circuits detected.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the foreign material on the catheter tip could not be conclusively determined.
|