An unspecified issue was reported with the adc device in use with an iphone 12 mini phone with an ios operating system version 1731.Customer was unable to access their sensors as their freestyle librelink account was no detecting the sensor.As a result, the customer was unable to monitor glucose with sensor readings and reported feeling "weak and extremely tired and very shaky and collapsed" experiencing a loss of consciousness, requiring treatment of glucogel by their mother.There was no report of death or permanent impairment associated with this event.
|
The most probable root causes associated with this failure mode are software/data corruption or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The customer e reported sensors do not activate.With freestyle librelink application.The reported issue was unable to be replicated as the reported configuration of ios [ios 17.3.1] is not compatible with the freestyle librelink app.As the compatibility guide is provided to the customer and the incompatible configurations were used, this complaint is not confirmed to use.The device manufacturing date does not apply.The date entered in section h4 is the date abbott diabetes care became aware of the event.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|