MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH CARDIOHELP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH CARDIOHELP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number CARDIOHELP
Device Problems Communication or Transmission Problem (2896); Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/26/2024
Event Type  malfunction  
Manufacturer Narrative
A getinge service technician will investigate the affected cardiohelp.A follow-up medwatch will be submitted when additional information becomes available.
 
Event Description
It was initially reported that the device had a loose contact and could not be used.Subsequently, it was reported that the internal arterial pressure was -500.The failure occurred during treatment.The values could be improved by plugging and unplugging the hls cable.No harm to any person has been reported.Complaint id # (b)(4).
 
Event Description
Complaint id # (b)(4).
 
Manufacturer Narrative
It was initially reported that the device had a loose contact and could not be used.Subsequently, it was reported that the internal arterial pressure was -500.The values could be improved by plugging and unplugging the hls cable.The failure occurred during treatment and the cardiohelp was exchanged.A getinge field service technician (fst) was sent for investigation.The fst confirmed that the pressure failure was related to the corrosion within the female connectors of the sensor panel.The sensor panel was replaced.The fst performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.The log files of the reported cardiohelp device were reviewed and no pump stops were observed on the date of event.According to the fts, the root cause was contamination in the female connector of the sensor panel.A similar issue was already investigated by the getinge life cycle engineering.Deposit was detected on the cable socket during visual inspection.The most probable root cause was determined as wetting of the socket plane of the plug with salt-containing liquids (priming).A service bulletin (issue 95 / 21-05-04) was published may 2021 to make the users aware that the contacts of the plug connections must not come into contact with cleaning agents, disinfectants or priming liquid.According to the instruction for use of the involved disposables (hls set advanced 5.0 / 7.0, hit set advanced 5.0 / 7.0, v2.4, chapter 6.1 preparation and installation and quadrox-ir small adult / adult, chapter 7.2 priming the system) the pressure sensors have to be calibrated and checked before priming.Furthermore, the cardiohelp has a flow/bubble sensor and a venous probe to measure and control the blood flow and parameters.If the measured values are above high limit or below low limit of the set limits the system generates a visual and acoustical alarm.In the instruction for use (ifu) of the cardiohelp, chapter "connection the sensors", it is stated to ensure that the connected sensors are not defective and to not use if there is a visible damage.Additionally, in the ifu of the cardiohelp (chapter ¿cleaning and disinfection") the cables and the whole device should be cleaned after each use to remove soiling or residual blood.Furthermore, in the ifu chapter "connecting the sensors" it is stated that the sensors must be kept clean.According to the ifu of the cardiohelp system, chapter ¿connecting the sensors¿, any connections that are not in use shall be protected with the supplied protective caps.Additionally, in the ifu, chapter ¿check before every application¿, it is stated that the device shall only be used after every function test prior to use has been passed.Furthermore, in the ifu, chapter ¿surface cleaning and disinfecting after each use¿, it is stated that the device shall only be cleaned with a damp cloth without using chemical solvents or spilling anesthetics.The device shall not be sprayed with liquids.The review of the non-conformities has been performed on 2024-03-18 for the period of 2017-03-16 to 2024-02-26.It does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.The device was manufactured on 2017-03-16.Based on the results the reported failure ¿pressure reading issue due to contaminated sensor panel" could be confirmed.The customer will be informed about the results by the getinge sales and service unit.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOHELP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key18864384
MDR Text Key337192212
Report Number8010762-2024-00127
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K133598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCARDIOHELP
Device Catalogue Number701048012
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2017
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-