For the investigation the logfile was analyzed.The described ventilation failure could be retraced in the logfile.It was found during on-site investigation by the service technician that after reprocessing obviously the cleaning disk was not removed from the upper rolling membrane of the piston.Nevertheless, the self-test was passed as the cleaning disk did not cause any leakage or device malfunction.It is assumed that during operation its position changed for example due to movement of the membrane, which caused the ventilation failure.The device alarmed as required by the safety concept and ventilation was continued in man/spont mode.After removing of the foreign material on site the device was successfully tested according to manufacturer's specification.During the manual checklist of the atlan self-test, the correct assembly of the breathing system/rolling membrane etc.Should be checked and the components should also be checked during installation after reprocessing.Therefore, it can be concluded that the event was caused by not following the instructions as described in the ifu.This is the first complaint describing this kind of problem.
|