U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDLINE INDUSTRIES, LP; DBD-SEPTOPLASTY PACK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDLINE INDUSTRIES, LP; DBD-SEPTOPLASTY PACK Back to Search Results
Model Number DYNJ17645A
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/19/2019
Event Type  malfunction  
Manufacturer Narrative
It was reported that a control syringe component's line of spray "goes all over" instead of being a "controlled line of spray." at the time of the original report, the reporting facility confirmed that the reported problem/issue did not cause or contribute to any injuries.To date, no information has been received to indicate that a user or a patient experienced a death, serious injury, medical intervention, follow-up care, or other adverse health impact associated with the reported problem/issue.A sample was returned for evaluation and visual inspection identified no damage to the component.Functional testing was performed and no major abnormalities were observed with the flow of the component.In an abundance of caution, and in response to an fda 483 issued for cfn (b)(4) on 22-jan-2024, this medwatch is being filed for the reported problem/issue.If additional relevant information becomes available a supplemental medwatch will be filed.
 
Event Description
It was reported that a control syringe component's line of spray "goes all over.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
DBD-SEPTOPLASTY PACK
Manufacturer (Section D)
MEDLINE INDUSTRIES, LP
three lakes drive
northfield IL 60093 2753
Manufacturer Contact
nigel vilches
three lakes drive
northfield, IL 60093-2753
MDR Report Key18865276
MDR Text Key337207637
Report Number1423395-2024-00215
Device Sequence Number1
Product Code OGR
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDYNJ17645A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2019
Date Manufacturer Received07/19/2019
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-