MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VIGILANT X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VIGILANT X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G247
Device Problems Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Read Input Signal (1581); Under-Sensing (1661)
Patient Problem Shock from Patient Lead(s) (3162)
Event Date 02/23/2024
Event Type  Injury  
Event Description
It was reported that this patient with this biventricular (biv) implantable cardioverter defibrillator (icd) experienced atrial fibrillation (af) with rapid ventricular response (rvr) in multiple episodes.Anti-tachycardia pacing (atp) and shocks were provided which were noted to possibly have been from the af with rvr.Intermittent atrial undersensing was observed along with irregular ventricular to ventricular intervals.In one episode the atrial electrogram (egm) appeared almost flat with no atrial sensing.The shock channel morphology showed ventricular conduction.The p waves had been measuring higher than at implant.All lead measurements were within normal range.Technical services (ts) discussed programming optimization including adjusting the therapy zones, with the field representative who would discuss with the physician.This biv icd remains in service.No adverse patient effects were reported.
 
Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIGILANT X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18865280
MDR Text Key337202509
Report Number2124215-2024-14306
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589287
UDI-Public00802526589287
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberG247
Device Catalogue NumberG247
Device Lot Number297925
Was Device Available for Evaluation? No
Date Manufacturer Received02/25/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age75 YR
Patient SexMale
-
-