• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA III COLONOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA III COLONOVIDEOSCOPE Back to Search Results
Model Number CF-HQ190L
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/06/2024
Event Type  malfunction  
Event Description
It was reported that during reprocessing, the colonovideoscope tested positive for less than 1000 colony forming units (cfus) of escherichia coli, 1 cfu of micrococci luteus, 1 cfu of streptococcus and less than 50 cfus of citrobacter freundii.The suction channels and insufflation channel were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Manufacturer Narrative
The suspect device has not yet been returned to olympus for evaluation and the investigation is ongoing.The physical device evaluation has been completed.Prior to the device evaluation, the device was sent out for additional microbiological testing.The hygiene microbiological investigation report indicated the suction and insufflation channels of the scope were cultured and found less than 1000 colony forming units (cfus) of escherichia coli, 1 cfu of micrococci luteus, 1 cfu of streptococcus and less than 50 cfus of citrobacter freundii.The results obtained comply with the target level for an endoscope subjected to high level disinfection and rinsed with sterile water.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III COLONOVIDEOSCOPE
Type of Device
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18865380
MDR Text Key337624953
Report Number9610595-2024-05017
Device Sequence Number1
Product Code FDF
UDI-Device Identifier04953170305115
UDI-Public04953170305115
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCF-HQ190L
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/14/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-