MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GREENLIGHT MOXY FIBER OPTIC; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GREENLIGHT MOXY FIBER OPTIC; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0010-2400
Device Problems Loss of or Failure to Bond (1068); Break (1069); Power Problem (3010); Firing Problem (4011)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/17/2024
Event Type  malfunction  
Manufacturer Narrative
There was no device available for analysis; however, the reported complaint was confirmed by the media provided by the costumer.Media also showed that the fiber was burned.A device history record review confirmed the device met all manufacturing specifications, a risk review confirmed that the event is accounted for in the risk documentation, a labeling review was performed and ifu includes specific warnings/instructions related to "fiber beam blinking/flashing/pulsing, fiber cap/tip fiber rotates within metal cap, fiber cap/tip break, fiber stopped working".There is no indication that the device was not used in accordance with the ifu.Based on the information provided, all compiled information on this investigation determines that the most probable cause is unintended use error caused or contributed to event since the interaction between the user and device, or sample, caused or contributed to the error.
 
Event Description
It was reported that during a photoselective vaporization of the prostate (pvp) procedure to treat benign prostatic hyperplasia (bph), the fiber beam began to intermittently flicker, and the fiber tip was rotating inside the metal cap of the fiber.Upon removing the fiber, the glass cap had broken off.The fiber had stopped working.The fiber was checked, and was no longer working, so another fiber was used to complete the procedure.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT MOXY FIBER OPTIC
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18865953
MDR Text Key337218367
Report Number2124215-2024-14195
Device Sequence Number1
Product Code GEX
UDI-Device Identifier00878953005515
UDI-Public00878953005515
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K120870
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0010-2400
Device Catalogue Number0010-2400
Device Lot Number0031089644
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/17/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/21/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age53 YR
Patient SexMale
-
-