Model Number VH-4000
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/15/2024
Event Type  malfunction  
Event Description
The hospital reported that during an endoscopic vessel harvesting procedure, vasoview hemopro 2 heater wire tip broke.No parts fell into patient.The jaws were not open during insertion of the tool into the cannula.Per photo provided by complainant, the metal heating wire separated from the jaw.A new device was opened to complete the procedure.There was a procedural delay.There was no patient harm.
 
Manufacturer Narrative
(b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Manufacturer Narrative
Trackwise #: (b)(4).Updated sections: b-4,d-9, g-3, g-6, h-2,h-3, h-6, h-10.A lot history record review was completed for lots 3000355899, 3000363687, and 3000360579 the last 3 lots shipped to the account prior to the event/aware date.For lot # 3000355899- there was one ncmr , rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.For lot # 3000363687 and 3000360579.There were no ncmrs.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key18866515
MDR Text Key337226791
Report Number2242352-2024-00219
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567700406
UDI-Public00607567700406
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup
Report Date 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-4000
Device Catalogue NumberVH-4000
Was Device Available for Evaluation? No
Date Manufacturer Received04/25/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-