Catalog Number RBYLP0210
Device Problems Physical Resistance/Sticking (4012); Premature Separation (4045)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/13/2024
Event Type  malfunction  
Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
The patient was undergoing a coil embolization procedure in the gastroduodenal artery (gda) using ruby coil lps and a non-penumbra microcatheter.During the procedure, after advancing the ruby coil lp into the target location, the physician decided to retract the coil to reposition the microcatheter.However, while retracting the ruby coil lp, the physician experienced resistance and the ruby coil lp unintentionally detached inside the microcatheter.Therefore, the microcatheter containing the detached ruby coil lp was removed.The procedure was completed using a new ruby coil lp and a new microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL LP
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key18866785
MDR Text Key337754125
Report Number3005168196-2024-00077
Device Sequence Number1
Product Code HCG
UDI-Device Identifier00815948025998
UDI-Public815948025998
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K230284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberRBYLP0210
Device Lot NumberF00008829
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received02/13/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-