ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL0¿, 63 CM LENGTH, 8.5 F; Introducer, catheter Back to Search Results
Model Number 407451
Device Problem Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2024
Event Type  malfunction  
Event Description
During an atrial fibrillation procedure, an x-ray image revealed the needle perforated the sheath when performing the transseptal puncture.However, a guide could still be inserted in the patient's left atrium.An agilis was used to perform the procedure (which had been planned to perform the ablation) with no consequences to the patient.An ett (exercise tolerance test) was performed at the end of the procedure and did not show an effusion or any unexpected / abnormal element.
 
Manufacturer Narrative
The device was not returned.Two images were submitted for evaluation to product performance engineering; no device components were returned.The photos appeared to show a swartz braided introducer.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported sheath puncture remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL0¿, 63 CM LENGTH, 8.5 F
Type of Device
Introducer, catheter
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key18866820
MDR Text Key337230280
Report Number3005334138-2024-00086
Device Sequence Number1
Product Code DYB
UDI-Device Identifier05414734205757
UDI-Public05414734205757
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K052644
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number407451
Device Lot Number10121673
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/04/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/12/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-