COLOPLAST A/S ALTIS SINGLE INCISION SLING SYSTEM; FEMALE STRESS URINARY INCONTINENCE SURGICAL MESH-SLING, SYNTHETIC POLYMER Back to Search Results
Catalog Number 519650
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/12/2024
Event Type  malfunction  
Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report, and capa.No ncs nor capas were identified in veeva.There were several complaints identified against this lot; however, with various failure modes.Therefore, this is not considered a trend.Devices met specification prior to release.
 
Event Description
According to the available information, the anchor releases immediately without force during implantation.This error occurred 3 times in a row during the procedure utilizing 3 distinct devices from the same lot.No adverse effects to the patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTIS SINGLE INCISION SLING SYSTEM
Type of Device
FEMALE STRESS URINARY INCONTINENCE SURGICAL MESH-SLING, SYNTHETIC POLYMER
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key18867243
MDR Text Key337234588
Report Number2125050-2024-00380
Device Sequence Number1
Product Code PAH
UDI-Device Identifier05701780274807
UDI-Public5701780274807
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K121562
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number519650
Device Lot Number9066451_5196504022
Is the Reporter a Health Professional? Yes
Date Device Manufactured08/31/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial