Catalog Number 21-7002-24 |
Device Problems
Device Alarm System (1012); Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that during use, the device exhibited an alarm for ¿no disposable pump will not run.¿ the pump was restarted multiple times, would run briefly, and the alarm would return.The error was not resolved.The patient presented to the clinic and was issued a new pump.The device delivered fluorouracil.Volume to be infused 100ml; rate 50ml/24hrs; volume delivered 3.65ml; volume remaining 96.4ml.Per the reporter, there were no adverse patient effects.
|
|
Manufacturer Narrative
|
Lot number unknown; possible lots: 4406177, 4396089, 4387666.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
|
|
Manufacturer Narrative
|
D9: date returned to mfg (b)(6) 2024.One sample was received for evaluation.Visual inspection revealed no discrepancies.Functional testing was performed and no discrepancies were found.The reported issue was not confirmed.No lot number was provided; therefore, a history record review could not be conducted.
|
|
Search Alerts/Recalls
|