It was reported, the ultrasonic surgical device was giving a probe damage error when the user tried to activate it.Upon examination, the probe tip was noted to ¿wobble¿ from the base as if the blade might come off.A second device was retrieved, which exhibited the same problems.A third device also had the same error but this time, the probe tip fell off completely.The issue occurred during a therapeutic, laparoscopic myoma enucleation, which was prolonged by 15 to 20 minutes while the patient was under general anesthesia.A non-olympus laparoscopic sealer/divider was then used instead, and the procedure was completed as open surgery.The patient was subsequently admitted to the ward due to the change to open surgery.There were no reports of patient harm.This report is for the first device.
|